Grip Sterilization:
We highly recommend CaviCide™ Cleaning and disinfectant solution & Madacide-FD or other solutions with similar compound.
The grip is compatible with autoclaves, ozone sterilizers and cold sterilization methods in solutions with non-aggressive acidity or alkalinity.
DO NOT autoclave or use dry heat sterilizers on the body of the pen! The motor, coating and mechanism will become unusable.
DO NOT let the grip contact any dirty surfaces during and after the process of sterilization.
DO NOT use dry heat sterilizer and do NOT use alkaline or acid cleaning agents with strong bases (high or low pH). Doing so will damage the grips and warranty will be void.
Cleaning:
Wipe the body of the pen, the grip and the cavity with the pushing bar inside using napkins and cotton buds with alcohol-based disinfection and sterilization solutions (alcohol evaporates quickly and won’t expose components of the pen to corrosion).
DO NOT soak neither body of the pen nor the grip in disinfection and sterilization solutions. Do not spray them over the pen either.
Cleaning & Grip Sterilization